Labomed Inc.


Double Beam Spectrophotometer

UVD-3200A

8 Automatic Cellholder, Variable Bandwidth 0.5/1/2/4/5

 

 

UVD-3200A Double Beam Spectrophotometer 8 Automatic Cellholder, Variable Bandwidth 0.5/1/2/4/5nm. UV-VIS Double Beam Spectrophotometer Scanning System with memory Photometric Range -0.3~3A and 8 Automatic Cellholder.

The UV-Vis with a linear range of 3A. Double Beam Spectrophotometer has a Photometric Range of -0.3~3A.

The Double Beam Spectrophotometer with Photometric Range -0.3~3A and Spectral Bandwidth 0.5/1/2/4/5 is designed to meet high requirement for precision measurement in the research and production of organic chemistry, biochemistry, medical testing, food testing, environmental protection, water testing industry, etc. The latest ARM system and long optical system ensure high accuracy and good stability of the instrument. They are the best choice of high quality spectrophotometer.

 

Labomed, Inc. is certified by ISO 9001-2013, has CE Conformity and is FDA Licensed

 

 

Features

Top of the Page

  • 7 inches TFT color screen and windows graphic interface.
  • Powerful functions like Photometric measurement, Quantitative measurement,
    Kinetics, Spectrum scan, DNA\Proteintest, multi-wavelength test, etc
  • In-house massive memory is capable of saving up to 1024M for test data &
    working curves.
  • Supports USB storage. The USB port can be used for data transfer, which is
    easily exported to Excel for further processing, analysis and storage.
  • Socket type deuterium and tungsten lamp can make lamps switching without
    optics debugging and easy to be replaced.
  • Large sample chamber can accommodate 5-100mm cuvettes of all kinds.
  • Extensive accessories are optional, such as auto 8-cell holder, film holder,
    peltier/sipper system, 21 CFR compliant software, built-in thermal printer, etc.

Technical Specifications

Top of the Page
Optical System Double Beam, Grating 1200 lines/mm
Wavelength Range 190-1100 nm
Spectral  Bandwidth 2nm
Wavelength Accuracy ±0.1nm@656.1nm, ±0.3nm@all
Wavelength Repeatability <=0.1nm
Photometric Accuracy 0.2%T(0~100%T), ±0.002A(0-0.5A), ±0.004A(0.5-1A)
Photometric Repeatability ≤0.15%T (0-100%T), 0.001A(0-0.5A), 0.002A(0.5-1A)
Photometric Range 0-200%T, -0.3~3A, 0-9999C (0-9999F)
Stray Light <=0.03%T@220nm, 360nm
Stability ±0.0003A/h @500 nm
Baseline Flatness ±0.001A
Noise 0.0005A@500nm
Work Mode T, A, C, E
Scanning Speed Hi, Med, Low (Max. 3000nm/min)
Wavelength Setting Auto
Display 7" TFT Color Screen
Light Source Imported Deuterium &Tungsten lamp
Detector Imported Silicon Photodiode
Cuvette Holder 10mm manual 4-cell holder
Output USB drive, USB host, RS232
Power AC 220V/50Hz or AC 110V/60Hz
Shipping Size 810*660*390mm
Gross Weight 27kg
Standard Accessaries 10mm glass cuvette x 4, 10mm quartz cuvette x 2, Power cord, User's Manual , PC software
   
   

 

 

 